General Description

Based on the rugged Humirel HTS2010 humidity sensor the HTM2500 is a dedicated humidity and temperature transducer designed for applications where a reliable and accurate measurement is needed.




 


Typical Applications

 

Building science research
  1. Building envelope cavity validation
  2. Indoor and outdoor air quality analysis
Restoration
  1. Verification of equipment status
  2. Drying progress
Industrial Applications
  1. Process control
  2. Hygrostat

Features

  1. Hermetic Housing
  2. Humidity calibrated within ±2% @55% RH
  3. Integrated Thermistor – MF52 pearl-shaped precision NTC thermistor (cable version)
  4. Small size
  5. Compatible with SMT Mobile and Industrial WiDAQ
  6. Full interchangeability
  7. High reliability and long term stability
  8. Not affected by water immersion
  9. Instantaneous de-saturation after long period in saturation phase
  10. Fast response time suitable for low voltage wireless applications.
  11. High resistance to chemicals
  12. Unique solid polymer structure

Specs

Electrical Characteristics

Operating Voltage
(with SMT WiDAQ)

0V to 5VDC

Sensing Element

HTM2500

Environmental

Operating Temperature

-30° to 70°C

Operating Humidity Range

0% to 100% RH

Storage Temperature

-40° to 85°C

Storage Humidity

0% to 100% RH

Humidity Characteristics

Humidity Measuring Range

1% to 99% RH

Relative Humidity Accuracy (10 to 95% RH)

±3 to ±5 %RH

WiDAQ Supply Voltage

5VDC

Current Consumption

0.4mA

Temperature Coefficient (10°C to 50°C)

±0.1 %RH/°C

Average Sensitivity from 33% to 75% RH

+25 mV/%RH

Recovery time after 150 hours of condensation

10 seconds

Humidity Hysteresis

±1.5 %RH

Long term stability

±0.5 %RH/yr

Time Constant (at 63% signal, static) 33% to 76% RH

5 seconds

Thermistor Characteristics

Rated Resistance R25

10 to 250 KΩ

B Value (25/50°C)

4150K

Operating Temperature

-55° to 125°C

Tolerance

1%

Approvals/Regulatory

Passed Meas-France qualification process

Vibration, shock, storage temperatures, high temperature and humidity and ESD.

Chemical conditions tested

Salt atmosphere, SO2, NOx, NO, CO, Softener, Soap, Toluene, acids (H2SO4, HNO3, HCI), HMDS, insecticide, cigarette smoke.  HTM255 is not light sensitive.

 

Connectivity

 

HTM2500 to Audio Jack

Insert RH Sensor in WiDAQ AUX Port

 

 

 


 HTM2500 to 8 Channel WiDAQ

HTM2500 Wiring to 8 Resistance 2 Voltage WiDAQ

Building Analytics Login

News

  • SMT introduces the SMT-A2 more...
  • SMT introduces the Building Intelligence Gateway Version 1.0 more...
  • SMT receives 2011 Insurance Canada Technology Award more...
  • SMT is ISO9001 Registered more...